The MCNP5 Monte Carlo code was used to simulate the dosimetry of an M‐19 iridium‐192 high dose rate brachytherapy source in both air/vacuum and water environments with the in‐air photon spectrum filtered to remove low‐energy photons below δ = 10 keV . Dosimetric data was organized into an away‐along table and was used to derive the updated AAPM Task Group Report No. 43 (TG‐43U1) parameters includingS K ,D ̇ ( r , θ ) , Λ ,g L ( r ) , F ( r , θ ) ,φ an ( r ) , andφ ¯ anand their respective statistical uncertainties.