WE‐C‐330A‐04: Effect of Projection Angles Used in Multi‐View Reconstruction (MVR) Using Images From a Microangiographic (MA) Detector and An Image‐Intensifier (II) System
Author(s) -
Patel V,
Kuhls A,
Noël P,
Walczak A,
Ionita C,
Chityala R,
Tranquebar R,
Rangwala H,
Kasodekar S,
Yadava G,
Hoffmann K,
Bednarek D,
Rudin S
Publication year - 2006
Publication title -
medical physics
Language(s) - English
Resource type - Journals
SCImago Journal Rank - 1.473
H-Index - 180
eISSN - 2473-4209
pISSN - 0094-2405
DOI - 10.1118/1.2241680
Subject(s) - imaging phantom , standard deviation , projection (relational algebra) , pixel , image intensifier , detector , cone beam computed tomography , optics , image resolution , flat panel detector , root mean square , iterative reconstruction , physics , nuclear medicine , mathematics , artificial intelligence , computer science , computed tomography , medicine , radiology , statistics , algorithm , quantum mechanics
Purpose: The sensitivity of a new 3D Multi‐View Reconstruction (MVR) angiography technique to the projection angles used is evaluated by comparing 3D centerlines calculated from combinations of three projections acquired from two imaging systems with that from micro‐Cone Beam CT (μCBCT), which is taken as truth. Method and Materials: A 3D centerline of a contrast‐filled carotid vessel phantom was reconstructed from image data acquired using a custom‐made μCBCT system with a microangiographic (MA) detector (45 μm pixels, 4.5 cm field‐of‐view (FOV)). Projection images of the same phantom were also acquired using the MA and an image intensifier (II) detector system (120 μm pixels, 4.5 in FOV) on a C‐arm x‐ray unit. The MVR technique was used to compute 3D centerlines for 12 combinations of projection angles. Each 3D MVR centerline was aligned with the μCBCT “true” 3D centerline using a Procrustes technique, and a root‐mean‐square (RMS) deviation was calculated. Results: The average RMS deviation for the MA‐MVR centerlines is 25 μm with a standard deviation of 3 μm over the 12 different projection‐angle combinations, whereas the average RMS deviation for the II‐MVR centerlines is 41 μm with a standard deviation of 4 μm over these same combinations. The RMS deviation as a percent of the internal vessel diameter, 0.75 mm, is 3.3% for the MA and 5.5% for the II and appears to be independent of view selection. Conclusion: For the MVR technique, the improved resolution of the MA resulted in improved centerline determination compared to the II system. For both detectors, the selection of a particular projection set had little effect on the RMS centerline deviation. The low RMS deviations for both detectors indicate that the MVR technique can provide accurate 3D centerlines. (Partial support from NIH Grants R01‐NS43924, R01‐EB02873, R01‐HL52567, R01‐EB02916, and Toshiba Medical Systems Corporation).

This content is not available in your region!

Continue researching here.

Having issues? You can contact us here