SU‐FF‐T‐318: Measured Versus Calculated Dynamic Wedge Dose Distribution Using Anisotropic Analytic Algorithm (AAA) and Pencil Beam Convolution (PBC) Algorithms
Author(s) -
Caprile P,
Venencia C,
Besa P
Publication year - 2006
Publication title -
medical physics
Language(s) - English
Resource type - Journals
SCImago Journal Rank - 1.473
H-Index - 180
eISSN - 2473-4209
pISSN - 0094-2405
DOI - 10.1118/1.2241236
Subject(s) - imaging phantom , algorithm , wedge (geometry) , linear particle accelerator , nuclear medicine , physics , mathematics , beam (structure) , geometry , optics , medicine
Purpose: The purpose of this work is to compare measured and calculated 2D EDW dose distributions using AAA and PBC calculations algorithms. Method and Materials: A 6MV photon beam from a Clinac 2300C/D Linac equipped with 7 EDW was used. Dose distributions were calculated for square symmetric fields by a Varian Eclipse v7.3.1 3DTPS with AAA and PBC algorithms. EDW were commissioning in the TPS. Dose distributions were measured using Kodak EDR2 films in a perpendicular configuration at 5cm depth in a solid water phantom for 4×4, 10×10, 15×15 and 20×20 cm field size settings. TPS calculations were performed for the same conditions as measurements, using a phantom with the same geometry. Calculation grid was 2.5×2.5 mm for PBC and 1×1 mm for AAA. Dose distributions were compared using RIT v4.3 software with gamma evaluation using 3% dose variation and 3mm DTA criteria. Results: Measured versus calculated percentage depth doses and wedge factors agree within 1%. Profiles in the non wedged direction, exhibit variations lower than 2% of dose or 2 mm DTA. In wedged direction both algorithms reproduces the measured profiles with deviations smaller than 2% of the normalization dose except in the penumbra region. AAA algorithm models better the penumbra region. EDW with angle lower than 45°, both algorithms model the distributions within the acceptance criteria. Some differences could be seen in the corners of the beams. For EDW with 60° angle, the percentage of pixels that not passes the acceptance criteria using AAA is acceptable, but using the PBC algorithm the percentage must be considered (reaching 28.5% of the distribution), representing regions of disagreement that may be of clinical interest. Conclusions: AAA algorithm models EDW dose distributions better than PBC. Differences between both algorithms increase with field size.

This content is not available in your region!

Continue researching here.

Having issues? You can contact us here