SU‐FF‐T‐158: Determination of the Energy Spectrum of Bremsstrahlung of Linear Accelerations by Monte‐Carlo Calculations and An Analysis of Depth Dose Curves
Author(s) -
Ulmer W
Publication year - 2006
Publication title -
medical physics
Language(s) - English
Resource type - Journals
SCImago Journal Rank - 1.473
H-Index - 180
eISSN - 2473-4209
pISSN - 0094-2405
DOI - 10.1118/1.2241082
Subject(s) - monte carlo method , bremsstrahlung , kernel (algebra) , computational physics , laplace transform , physics , mathematics , photon , mathematical analysis , optics , statistics , combinatorics
Purpose: The knowledge of the bremsstrahlung spectrum of a linear accelerator (6 MV up to 18 MV) represents the physical base of therapy planning systems. It is obtained by Monte‐Carlo calculations by taking account of the complete beam‐line (geometry/materials). It is also obtained by an analysis of depth dose curves of small fields via Laplace transform, if the electron energy E el is known (upper limit of photon energy). Method and Materials: Monte‐Carlo calculations with EGSnrc with respect to the beam‐line of 6/18 MV (Varian 2300 C/D) and 6 MV (Varian 600C) have been performed for the AAA algorithm (Eclipse). The analysis of absorption depth dose curves in water (3×3 cm 2 / 4×4 cm 2 fields) of the mentioned machines has been performed for the central ray and tilted rays with the Laplace transform of a power sequence of Poisson distributions, which provide the reproducing kernel and the energy spectrum. The scatter profiles have been removed by a deconvolution. Results: Monte‐Carlo calculations and the Laplace transform method are in good agreement (standard deviation: ca. 1%) in the above cases. The formal integration of the reproducing kernel provides an analytical spectral distribution function f ( E ) = ( 1 − exp ( − α ⋅ E / E el) ) ⋅ exp ( − β E 2/   E el 2) ⋅ ( 1 − E / E el ) qThe parameter α and q depend on the radial distance of tilted rays from the central ray at surface. It is also used to fit fluctuations of Monte‐Carlo calculations. Conclusion: The Laplace transform method even works, if only E el , the measured depth dose curves and profiles of small field sizes are known. A comparison of 6 MV (2300 C/D) with 6 MV (600C) shows that the spectral distribution of the latter case rather corresponds to the 10 MV mode of a 2300 C/D machine. This results from the Pb‐alloy of the flattening filter, whereas the 6/10 MV modes of the 2300 C/D use a Cu‐filter.

This content is not available in your region!

Continue researching here.

Having issues? You can contact us here
Accelerating Research

Address

John Eccles House
Robert Robinson Avenue,
Oxford Science Park, Oxford
OX4 4GP, United Kingdom