Author(s) -
Åslund Magnus,
Cederström Björn,
Lundqvist Mats,
Danielsson Mats
Publication year - 2006
Publication title -
medical physics
Language(s) - English
Resource type - Journals
SCImago Journal Rank - 1.473
H-Index - 180
eISSN - 2473-4209
pISSN - 0094-2405
DOI - 10.1118/1.2179122
Subject(s) - mammography , optics , digital mammography , detector , detective quantum efficiency , materials science , digital radiography , physics , nuclear medicine , image quality , radiography , medicine , nuclear physics , breast cancer , computer science , cancer , artificial intelligence , image (mathematics)
The scatter to primary ratio (SPR) was measured on a scanning multislit full‐field digital mammography system for different thickness of breast equivalent material and different tube voltages. Scatter within the detector was measured separately and was found to be the major source of scatter in the assembly. Measured total SPRs below 6% are reported for breast range 3 – 7 cm . The performance of the multislit assembly is compared to other imaging geometries with different scatter rejection schemes by using the scatter detective quantum efficiency.