SU‐FF‐T‐294: Monte Carlo Simulations of the Dosimetric Characteristics of a New Multileaf Collimator
Author(s) -
Tacke M,
Szymanowski H,
Schulze C,
Nuss S,
Wehrwein E,
Leidenberger S,
Oelfke U
Publication year - 2005
Publication title -
medical physics
Language(s) - English
Resource type - Journals
SCImago Journal Rank - 1.473
H-Index - 180
eISSN - 2473-4209
pISSN - 0094-2405
DOI - 10.1118/1.1998023
Subject(s) - multileaf collimator , monte carlo method , collimator , electromagnetic shielding , materials science , dosimetry , optics , computational physics , physics , nuclear medicine , linear particle accelerator , beam (structure) , mathematics , medicine , statistics , composite material
Purpose: The aim of the work was to investigate the dosimetric characteristics of a new multileaf collimator (160MLC™, Siemens) with the help of Monte Carlo (MC) simulations during the design phase. Method and Materials: The MLC was implemented in the MC code Geant4. For the simulation of the 6 MV treatment beam an experimentally validated phase space and a virtual source model were used. For the simulation of the geometry in Geant4 the jaws and the two leaf packages were implemented with the help of CAD data. First, transmission values for different tungsten sinters were extracted using the simulation codes Geant4 and BEAMnrc and compared to experimental measurements. In a second step, high resolution simulations were performed to detect the leakage at depth of maximum dose. The 20%–80% penumbra along the leaf travel direction was determined for different 10×10 cm 2 fields shifted along the x‐axis. The simulated results were compared with measured data obtained with a prototype. Results: The simulation of the transmission values for different tungsten sinters showed a good agreement with the experimental measurements (within 2.0%). This gave an accurate estimation of the absorption coefficient for various leaf materials. Simulations with varying source sizes showed that the leakage and the penumbra depended very much on this parameter: e.g. source sizes of 2 mm and 4 mm result in the interleaf leakages below 0.3% and 0.75% respectively. The results for the leakage and the penumbra are in good agreement with the measurements. Conclusion: This study showed that Geant4 is appropriate for the investigation of the dosimetric characteristics of a multileaf collimator. In particular we could quantify the leakage and the penumbra and evaluate the influence of the beam parameters such as the virtual source size. Conflict of Interest: Research supported by Siemens Oncology Care Systems.