SU‐FF‐T‐146: How Monthly QA Affect the IMRT QA Result
Author(s) -
Grimm S,
Zhou F,
Grimm J,
Baker A
Publication year - 2005
Publication title -
medical physics
Language(s) - English
Resource type - Journals
SCImago Journal Rank - 1.473
H-Index - 180
eISSN - 2473-4209
pISSN - 0094-2405
DOI - 10.1118/1.1997817
Subject(s) - isocenter , imaging phantom , field size , nuclear medicine , mathematics , range (aeronautics) , percentage depth dose curve , dosimetry , beam (structure) , field (mathematics) , monitor unit , optics , physics , ionization chamber , medicine , materials science , ion , quantum mechanics , pure mathematics , composite material , ionization
Purpose: According to TG 40, film demonstrates the width and length of field between 9.8cm to 10.2 cm are considered acceptable for a 10×10 cm field size. This work is to demonstrate under these two accepted scenarios, how IMRT QA results are affected. Method and Materials: Five IMRT QA plans are generated with each gantry angle set to the nominal angle (down onto phantom surface) and isocenter located at 100cm SAD, depth 10 cm. A 2D diode array MAPCHECK was set up and radiated according to the QA plan. These five QA plans were repeatedly radiated on four different days. The first exposure was done prior to the field adjustment with field size 9.8×9.9 cm for a 10×10 cm field size; the subsequent three measurements were done in three different days after the field adjustment to 10.12×9.9cm. Absolute Measured dose is compared to planned absolute dose for each detector. We also compared the measurement data with the different beam modeling sigma parameter. Results: Prior to the field adjustment, the passing rate for these five QA plans ranged from 42.1% to 78.3 % with 3% Dose difference and 2mm of DTA. The passing rate increased from 64.5% to 91.8% with 3% Dose difference and 3 mm of DTA. Table I present these comparisons. After the field adjustment, the passing rate range from 91.5% to 98.3 % with 3% Dose difference and 2mm of DTA. Table II, III and IV present the comparisons. Conclusion: Even though the field size for 9.8×9.9 cm is acceptable for a 10×10 field size for monthly QA, the consistent reduction of the field size (2mm) all the way to 1×1 dramatically impacts IMRT QA results much more so than the adjusting the sigma parameter on the beam modeling data.

This content is not available in your region!

Continue researching here.

Having issues? You can contact us here