A phantom study on the positioning accuracy of the Novalis Body system
Author(s) -
Yan Hui,
Yin FangFang,
Kim Jae Ho
Publication year - 2003
Publication title -
medical physics
Language(s) - English
Resource type - Journals
SCImago Journal Rank - 1.473
H-Index - 180
eISSN - 2473-4209
pISSN - 0094-2405
DOI - 10.1118/1.1626122
Subject(s) - isocenter , imaging phantom , positioning system , accuracy and precision , rotation (mathematics) , nuclear medicine , computer science , physics , optics , computer vision , medicine , acoustics , quantum mechanics , node (physics)
A phantom study was conducted to investigate inherent positioning accuracy of an image‐guided patient positioning system—the Novalis Body system for three‐dimensional (3‐D) conformal radiotherapy. This positioning system consists of two infrared (IR) cameras and one video camera and two kV x‐ray imaging devices. The initial patient setup was guided by the IR camera system and the target localization was accomplished using the kV x‐ray imaging system. In this study, the IR marker shift and phantom rotation were simulated and their effects on the positioning accuracy were examined by a Rando phantom. The effects of CT slice thickness and treatment sites on the positioning accuracy were tested. In addition, the internal target shift was simulated and its effect on the positioning accuracy was examined by a water tank. With the application of the Novalis Body system, the positioning error of the planned isocenter was significantly reduced. The experimental results with the simulated IR marker shifts indicated that the positioning errors of the planned isocenter were 0.6±0.3, 0.5±0.2, and 0.7±0.2 mm along the lateral, longitudinal, and vertical axes, respectively. The experimental results with the simulated phantom rotations indicated that the positioning errors of the planned isocenter were 0.6±0.3, 0.7±0.2, and 0.5±0.2 mm along the three axes, respectively. The experimental results with the simulated target shifts indicated that the positioning errors of the planned isocenter were 0.6±0.3, 0.7±0.2, and 0.5±0.2 mm along the three axes, respectively. On average, the positioning accuracy of 1 mm for the planned isocenter was achieved using the Novalis Body system.

This content is not available in your region!

Continue researching here.

Having issues? You can contact us here