Direct conversion detectors: The effect of incomplete charge collection on detective quantum efficiency
Author(s) -
Mainprize James G.,
Hunt Dylan C.,
Yaffe Martin J.
Publication year - 2002
Publication title -
medical physics
Language(s) - English
Resource type - Journals
SCImago Journal Rank - 1.473
H-Index - 180
eISSN - 2473-4209
pISSN - 0094-2405
DOI - 10.1118/1.1477235
Subject(s) - detective quantum efficiency , detector , noise (video) , optics , physics , quantum efficiency , quantum noise , x ray detector , quantum , image quality , computer science , quantum mechanics , artificial intelligence , image (mathematics)
Direct conversion detectors offer the potential for very high resolution and high quantum efficiency for x‐ray imaging, however, variations in signal can arise due to incomplete charge collection. A charge transport model was developed to describe the signal and noise resulting from incomplete charge collection. This signal to noise ratio (SNR) reduction was incorporated into the cascaded systems model for a simple x‐ray detector. A new excess noise factor, A c(termed the collection noise factor) is introduced to describe the reduction in detective quantum efficiency (DQE). The DQE is proportional to the product of the quantum efficiency and the collection noise factor. If the trapping cross sections for electrons and holes are very different, and if the detector is biased improperly, the collection noise factor can drop to as low as 50%. In addition, the signal loss due to incomplete charge collection will reduce the DQE in the presence of added noise. Because of this, the DQE generally does not continue to improve with greater detector thickness. The collection noise factor and DQE are predicted for CdZnTe, PbI 2 , and Se. The optimization of detector thickness should be based not only on quantum efficiency but also on the charge collection statistics, which are influenced by bias field and polarity.