Effect of Viscoelasticity on Arterial-Like Pulsatile Flow Dynamics and Energy
Author(s) -
Winston H. Elliott,
Dongjie Guo,
Gruschen Veldtman,
Wei Tan
Publication year - 2019
Publication title -
journal of biomechanical engineering
Language(s) - English
Resource type - Journals
SCImago Journal Rank - 0.546
H-Index - 126
eISSN - 1528-8951
pISSN - 0148-0731
DOI - 10.1115/1.4044877
Subject(s) - pulsatile flow , viscoelasticity , materials science , hemodynamics , cardiac cycle , shear stress , electrical conduit , mechanics , distension , biomedical engineering , lumen (anatomy) , cardiology , medicine , composite material , physics , mechanical engineering , engineering
Time-dependent arterial wall property is an important but difficult topic in vascular mechanics. Hysteresis, which appears during the measurement of arterial pressure-diameter relationship through a cardiac cycle, has been used to indicate time-dependent mechanics of arteries. However, the cause-effect relationship between viscoelastic (VE) properties of the arterial wall and hemodynamics, particularly the viscous contribution to hemodynamics, remains challenging. Herein, we show direct comparisons between elastic (E) (loss/storage < 0.1) and highly viscoelastic (loss/storage > 0.45) conduit structures with arterial-like compliance, in terms of their capability of altering pulsatile flow, wall shear, and energy level. Conduits were made from varying ratio of vinyl- and methyl-terminated poly(dimethylsiloxane) and were fit in a mimetic circulatory system measuring volumetric flow, pressure, and strain. Results indicated that when compared to elastic conduits, viscoelastic conduits attenuated lumen distension waveforms, producing an average of 11% greater cross-sectional area throughout a mimetic cardiac cycle. In response to such changes in lumen diameter strain, pressure and volumetric flow waves in viscoelastic conduits decreased by 3.9% and 6%, respectively, in the peak-to-peak amplitude. Importantly, the pulsatile waveforms for both diameter strain and volumetric flow demonstrated greater temporal alignment in viscoelastic conduits due to pulsation attenuation, resulting in 25% decrease in the oscillation of wall shear stress (WSS). We hope these findings may be used to further examine time-dependent arterial properties in disease prognosis and progression, as well as their use in vascular graft design.