Evaluation of a point‐of‐care blood glucose monitor in healthy goats
Author(s) -
Quandt Jane E.,
Barletta Michele,
Cornell Karen K.,
Giguère Steeve,
Hofmeister Erik H.
Publication year - 2018
Publication title -
journal of veterinary emergency and critical care
Language(s) - English
Resource type - Journals
SCImago Journal Rank - 0.886
H-Index - 47
eISSN - 1476-4431
pISSN - 1479-3261
DOI - 10.1111/vec.12686
Subject(s) - venipuncture , medicine , limits of agreement , butorphanol , premedication , xylazine , concordance , anesthesia , venous blood , point of care , concordance correlation coefficient , zoology , nuclear medicine , ketamine , pathology , mathematics , biology , statistics
Objective To assess agreement between a point‐of‐care glucometer (POCG) and a laboratory chemistry analyzer for blood glucose measurements in goats. Design Prospective study. Setting University teaching hospital. Animals Eighteen healthy adult goats. Investigations Whole blood samples were obtained via jugular venipuncture prior to premedication with xylazine and butorphanol (T0), following premedication (T20), and after 1 hour of inhalant anesthesia (T60). Each sample was tested with a POCG and a laboratory analyzer (HITA). Agreement was assessed using concordance correlation coefficients and calculation of bias and 95% limits of agreement. Measurements and Main Results Mean blood glucose concentration at T0 was 3.9 ± 0.6 mmol/L (70 ± 10 mg/dL; POCG) and 2.9 ± 0.4 mmol/dL (53 ± 8 mg/dL; HITA). Glucose concentrations at T20 were 6.7 ± 2.4 mmol/L (121 ± 43 mg/dL) and 5.4 ± 2.1 mmol/L (97 ± 37 mg/dL) and at T60 were 5.7 ± 1.7 mmol/L (102 ± 31 mg/dL) and 4.7 ± 1.3 mmol/L (85 ± 24 mg/dL) when measured with the POCG and HITA, respectively. The POCG overestimated blood glucose compared to the HITA. The bias ± SD was 1.08 ± 0.53 mmol/L (19.4 ± 9.5 mg/dL) (95% LOA 0.04 to 2.11 mmol/L [0.7 to 38.0 mg/dL]) and the concordance correlation coefficient was 0.82. After correcting the results of the POCG using a mixed‐effects linear model, the bias was 0.0 ± 0.38 mmol/L (0.0 ± 6.8 mg/dL) (95% LOA ± 0.74 mmol/L [± 13.4 mg/dL]) and the concordance correlation coefficient was 0.98. Conclusions The POCG overestimated blood glucose concentrations in goats, compared to the HITA, but when the POCG concentrations were corrected, the agreement was excellent.

This content is not available in your region!

Continue researching here.

Having issues? You can contact us here