Validation of the handheld Lactate‐Pro analyzer for measurement of blood L‐lactate concentration in cattle
Author(s) -
Buczinski Sébastien,
Doré Elizabeth,
Boulay Guillaume,
Francoz David
Publication year - 2014
Publication title -
veterinary clinical pathology
Language(s) - English
Resource type - Journals
SCImago Journal Rank - 0.537
H-Index - 51
eISSN - 1939-165X
pISSN - 0275-6382
DOI - 10.1111/vcp.12185
Subject(s) - coefficient of variation , medicine , reproducibility , cutoff , gold standard (test) , confidence interval , nuclear medicine , spectrum analyzer , blood lactate , chromatography , chemistry , heart rate , blood pressure , physics , quantum mechanics , electrical engineering , engineering
Background Blood L‐lactate concentration ( LAC ) can be used for various diagnostic purposes in cattle. As multiple handheld analyzers for LAC exist, it is important to validate their use in cattle in comparison with reference laboratory blood analyzers. Objectives The objectives of this study were to validate the handheld Lactate Pro meter (LacP) including reproducibility, and compare the measurements with the StatProfile (StatP) as a gold standard. In addition, diagnostic sensitivity and specificity, and the impact of HCT on LAC measured by both analyzers were assessed. Methods A cohort of 64 cattle with acute medical and surgical conditions was studied. Whole blood samples in heparin lithium tubes were analyzed upon arrival with both StatP and LacP. Twenty‐three samples were immediately retested to assess intra‐assay coefficient of variation ( CV ). The HCT values were also recorded. Results The LAC using LacP was highly correlated with the StatP ( r  = 0.9736 [95% confidence interval [CI]: 0.9562–0.9841]). The LacP underestimated LAC (mean difference:−0.9 mmol/L, 95% CI :−3.1 mmol/L to 1.3 mmol/L). The intra‐assay CV was excellent (4.77%). No significant correlation was observed between LacP or StatP and HCT ( P  = .39 and .09, respectively). Sensitivity and specificity for LacP were 91.7% (95% CI : 76.4–97.8%) and 100% (83.4–100%, cutoff of 4 mmol/L), and 78.6% (58.5–90.9%) and 100% (87.0–100%, cutoff of 6 mmol/L). Conclusions The LacP handheld lactate meter can be used safely and reliably cow‐side, although it underestimates LAC value when compared with a standard laboratory analyzer especially for LAC  ≥ 10.0 mmol/L. The LAC value was not influenced by HCT in this study.

This content is not available in your region!

Continue researching here.

Having issues? You can contact us here