Premium
Comparison of two apheresis systems for autologous stem cell collections in pediatric oncology patients
Author(s) -
EvenOr Ehud,
EdenWalker Alexandra,
Di Mola Maria,
McDougall Elizabeth,
Schechter Tal,
Ali Muhammad,
Svajger Gordana,
Gassas Adam,
Licht Christoph,
Krueger Joerg
Publication year - 2017
Publication title -
transfusion
Language(s) - English
Resource type - Journals
SCImago Journal Rank - 1.045
H-Index - 132
eISSN - 1537-2995
pISSN - 0041-1132
DOI - 10.1111/trf.13859
Subject(s) - medicine , apheresis , blood collection , demographics , cd34 , nuclear medicine , surgery , stem cell , urology , platelet , emergency medicine , demography , sociology , biology , genetics
BACKGROUND Peripheral stem cell collections can be challenging in the pediatric population and respective experience is limited. Since February 2015 our institution is utilizing the new Spectra Optia (Optia) apheresis device, which has replaced the former COBE Spectra (COBE) device. As a quality initiative we collected and compared collection efficiency (CE2) and other collection variables between the two devices. STUDY DESIGN AND METHODS In this retrospective study we collected and compared clinical, laboratory, and technical collection data from stem cell collection procedures done with the Optia and COBE devices. The collected data included patient demographics, precollection peripheral CD34+ cell counts, total CD34+ cells collected, complete blood count, electrolytes before and after collection, side effects attributed to the collection, total blood volumes processed (TBVs), collection times, and calculated CE2 and collection ratios. RESULTS Forty‐one collection procedures performed on 29 pediatric patients with the Optia device were compared to 41 collections performed on 27 patients with the COBE device. The TBVs through the Optia device were significantly smaller than the COBE (3.9 ± 0.2 × TBV vs. 5.5 ± 0.1 × TBV, respectively; p < 0.001), requiring significantly less anticoagulant and providing similar amounts of stem cells while collection times were significantly shorter (mean, 238 ± 9 min vs. 264 ± 9 min, respectively; p < 0.05). Collections on the Optia caused significantly smaller reductions of plasma calcium and magnesium. No significant side effects attributed to the procedure were noted. CONCLUSION Stem cell apheresis with the Optia device in children is safe and feasible with smaller blood volumes with shorter collection times.