Left ventricular inhibition during cardiac resynchronization caused by sensed left‐sided ventricular premature complexes
Author(s) -
Barold S. Serge,
Kucher Andreas
Publication year - 2020
Publication title -
pacing and clinical electrophysiology
Language(s) - English
Resource type - Journals
SCImago Journal Rank - 0.686
H-Index - 101
eISSN - 1540-8159
pISSN - 0147-8389
DOI - 10.1111/pace.13868
Subject(s) - medicine , cardiac resynchronization therapy , cardiology , ventricular function , cardiac function curve , qrs complex , heart failure , ejection fraction
A recently published case documented left ventricular (LV) inhibition of a Boston Scientific device by a premature complex (VPC) that was undetected by the right ventricular channel. We have observed a similar response in two patients with a BIOTRONIK cardiac resynchronization device also designed with LV sensing. A spurious response simulating that of the two true cases was also observed in a third patient with a defective LV lead which created isolated false signals. The responses of the BIOTRONIK devices were identical to that of the previously reported findings with the Boston Scientific device. The observations provide insight into the timing function of cardiac resynchronization devices designed with LV sensing.