Inappropriate disabling of an ICD noise‐detection algorithm in pacemaker‐dependent patients
Author(s) -
Welte Nicolas,
Gang Eli S.,
Swerdlow Charles D.,
Strik Marc,
Bordachar Pierre,
Ploux Sylvain
Publication year - 2019
Publication title -
pacing and clinical electrophysiology
Language(s) - English
Resource type - Journals
SCImago Journal Rank - 0.686
H-Index - 101
eISSN - 1540-8159
pISSN - 0147-8389
DOI - 10.1111/pace.13564
Subject(s) - medicine , ventricular fibrillation , cardiology , ventricular tachycardia , implantable cardioverter defibrillator , algorithm , computer science
SecureSense is an implantable cardioverter defibrillator algorithm that differentiates lead‐related oversensing from ventricular tachycardia/ventricular fibrillation by continuous comparison between the near‐field (NF) and the far‐field (FF) electrogram. If lead noise is identified, inappropriate therapy is withheld. Undersensing on the FF channel could result in inappropriate inhibition of life‐saving therapy. Thus, the device automatically switches SecureSense to passive mode if undersensing on the FF channel is suspected. We report here the first cases of inappropriate automatic SecureSense deactivation due to misdiagnosed FF undersensing in pacemaker‐dependent patients. Physicians should be aware that SecureSense does not withhold an inappropriate therapy for sustained oversensing in pacemaker‐dependent patients.