Compatibility of Radiofrequency Surgical Sponge Detection Technology with Cardiac Implantable Electronic Devices and Temporary Pacemakers
Author(s) -
SALCEDO JONATHAN D.,
PRETORIUS VICTOR G.,
HSU JONATHAN C.,
LALANI GAUTAM G.,
SCHRICKER AMIR A.,
HEBSUR SHRINIVAS M.,
M THOMAS J.,
HUNTER JESSICA A.,
LEWIS KATHRYN E.,
KRUMMEN DAVID E.,
FELD GREGORY K.,
BIRGERSDOTTERGREEN ULRIKA
Publication year - 2016
Publication title -
pacing and clinical electrophysiology
Language(s) - English
Resource type - Journals
SCImago Journal Rank - 0.686
H-Index - 101
eISSN - 1540-8159
pISSN - 0147-8389
DOI - 10.1111/pace.12938
Subject(s) - medicine , cardiac pacing , implantable cardioverter defibrillator , biomedical engineering , cardiology
Background Radiofrequency (RF) technology has improved detection of retained surgical sponges with a reported 100% sensitivity and specificity. However, the potential for interactions of the RF signals emitted by the detection system with cardiac implantable electronic devices (CIEDs) or temporary pacemakers may limit its use in those patients with these devices. This study investigated whether RF detection technology causes interference or clinically significant changes in the programmed settings of implanted pacemakers and defibrillators or temporary epicardial pacemakers. Methods Fifty patients who were scheduled either for CIED removal or placement of a temporary epicardial pacemaker (at the time of open heart surgery) were recruited for this study. Device settings and measurements from separate interrogations before and after scanning with the RF detection system were compared. For the temporary pacemakers, we observed for any changes in hemodynamics or signs of pacing interference. Results Twenty (40%) pacemakers, 20 (40%) implantable cardioverter defibrillators, and 10 (20%) temporary pacemakers were analyzed in this study. During scanning, no signal interference was detected in any permanent device, and there were no significant changes in programmed settings after scanning with the RF detection system. However, pacing inhibition was detected with temporary pacing systems when programmed to a synchronous mode (DDD). Conclusions RF detection technology can be safely used to scan for retained surgical sponges in patients with permanent CIEDs and temporary pacemakers set to asynchronous mode.

This content is not available in your region!

Continue researching here.

Having issues? You can contact us here