Effectiveness and Reliability of Selected Site Pacing for Avoidance of Phrenic Nerve Stimulation in CRT Patients with Quadripolar LV Leads: The EffaceQ Study
Author(s) -
OSWALD HANNO,
ASBACH STEFAN,
KÖBE JULIA,
WEGLAGE HEINRICH,
SCHULTEPITZKE BERNFRIED,
BRACHMANN JOHANNES
Publication year - 2015
Publication title -
pacing and clinical electrophysiology
Language(s) - English
Resource type - Journals
SCImago Journal Rank - 0.686
H-Index - 101
eISSN - 1540-8159
pISSN - 0147-8389
DOI - 10.1111/pace.12664
Subject(s) - medicine , coronary sinus , cardiology , cardiac resynchronization therapy , ambulatory , lead (geology) , heart failure , ejection fraction , geomorphology , geology
Background Cardiac resynchronization therapy (CRT) relies on sufficient left ventricular (LV) pacing with safety margin to phrenic nerve stimulation (PNS). Previous studies introduced LV vector reprogramming in bipolar coronary sinus leads to optimize LV pacing and avoid PNS. We investigated the efficacy and reliability of quadripolar leads in CRT. Methods The EffaceQ study enrolled 344 patients with de novo CRT implantation with a quadripolar LV lead in an observational, prospective multicenter study. The study was powered to demonstrate that in at least 90% of patients with an implanted quadripolar LV lead, a viable LV pacing configuration (LVPC) is available (primary end point: LV pacing threshold ≤2.5 V/0.5 ms, sufficient PNS margin). Results Quadripolar leads were successfully implanted in 96% of patients. A total of 278 of 299 (93.0%) patients with complete data met the criteria for viable LVPC. With the use of traditional LVPCs, a viable LVPC would have been available (268 of 299 patients; P = 0.002) in significantly fewer patients (89.6%). In any LVPC, PNS was inducible in 65.0% of patients and 22.6% of patients reported PNS during ambulatory 3‐month follow‐up. LVPC reprogramming was performed in 49.8% of patients. PNS inducibility decreased from distal to proximal electrodes, whereas LV pacing thresholds increased from distal to proximal. At prehospital discharge, 5.9 ± 2.8 viable LVPCs were observed, stable during follow‐up. The quadripolar electrode offered significantly more LVPC for LV optimization and PNS avoidance. Conclusion Quadripolar LV leads yield high numbers of patients with viable LVPCs and alternatives for noninvasive repositioning of LV pacing.