z-logo
Premium
Formation of Thermal Coagulum on Multielectrode Catheters during Phased Radiofrequency Energy Ablation of Persistent Atrial Fibrillation
Author(s) -
DEBRUYNE PHILIPPE,
ROSSENBACKER TOM,
VANKELECOM BART,
CHARLIER FILIP,
ROOSEN JOHN,
ECTOR BAVO,
JANSSENS LUC
Publication year - 2014
Publication title -
pacing and clinical electrophysiology
Language(s) - English
Resource type - Journals
SCImago Journal Rank - 0.686
H-Index - 101
eISSN - 1540-8159
pISSN - 0147-8389
DOI - 10.1111/pace.12256
Subject(s) - medicine , ablation , atrial fibrillation , catheter ablation , catheter , duty cycle , radiofrequency ablation , pulmonary vein , biomedical engineering , cardiology , surgery , electrical engineering , voltage , engineering
Background Radiofrequency ablation (RFA) can unfavorably cause coagulum on the ablation electrode. The aim of this study was to assess this phenomenon on three different multielectrode catheters used to treat persistent atrial fibrillation with duty‐cycled RFA. Methods and results Twenty‐six consecutive patients have been treated with the pulmonary vein ablation catheter (PVAC) and the multiarray ablation catheter (MAAC). In 13 patients, additional ablation with the multiarray septal catheter (MASC) has been performed. The multichannel RF generator GENius™ (Medtronic Inc., Minneapolis, MN, USA) independently delivered energy in a bipolar and unipolar mode (ratio of 4/1, 2/1, or 1/1) to any of the electrodes. Versions 14.2, 14.3, and 14.4 of the generator were used. Coagulum presence was determined postablation by careful visual inspection of the catheter electrodes. No coagulum formation was visualized on the PVACs. Coagulum formation was visualized in 59% of the electrodes of the MAACs using a 2/1 mode and the 14.2 software version versus 69% using the 14.4 version and a 2/1 mode (P = 0.7) versus 14% of the electrodes applying a 1/1 ratio and the 14.4 software version (P < 0.001). Conclusions Duty‐cycled RFA in 2/1 bipolar/unipolar ratio generates a substantial frequency of coagulum formation on the multielectrode catheters MAAC and MASC. The use of the 14.4 version of the software to drive the RF generator and the use of energy in the default 1/1 bipolar/unipolar ratio could significantly reduce the frequency of coagulum formation, but so far, could not completely overcome it. The PVAC did not form coagulum, regardless of generator version or energy ratio used.

This content is not available in your region!

Continue researching here.

Having issues? You can contact us here