Differential Lead Component Pulling as a Possible Mechanism of Inside‐Out Abrasion and Conductor Cable Externalization
Author(s) -
LAU ERNEST W.
Publication year - 2013
Publication title -
pacing and clinical electrophysiology
Language(s) - English
Resource type - Journals
SCImago Journal Rank - 0.686
H-Index - 101
eISSN - 1540-8159
pISSN - 0147-8389
DOI - 10.1111/pace.12185
Subject(s) - medicine , lead (geology) , abrasion (mechanical) , compression (physics) , cardiology , structural engineering , composite material , materials science , geology , geomorphology , engineering
Background Conductor cable externalization with protrusion (CCE*) is highly prevalent among the Riata 8F and ST 7F defibrillation (DF) leads and infrequently present in the QuickSite and the QuickFlex coronary sinus (CS) leads (St. Jude Medical, Sylmar, CA, USA). A model for CCE* based on differential lead component pulling and conjugate extension with reciprocal compression‐bending was developed. Extension of a proximal lead body segment by pectoral or cardiac movements causes reciprocal compression‐bending of a distal lead body segment mediated by inextensible conductor cables running down a lead body fixed at various points by fibrous adhesions. The “sawing” action of these cables under tension causes inside‐out abrasion of insulation leading to CCE*. Methods DF leads from different manufacturers and the QuickFlex and QuickFlex μ CS leads were subjected to simulated differential pulling. Results Restitution from differential pulling followed three patterns: complete, partial without escalation, and incomplete with escalation. Only the last pattern (only shown by the Riata 8F and ST 7F leads) was associated with an increased risk to CCE*. For CS leads, deformation concentrated on the more flexible segment when the lead body did not have a uniform construction. Conclusions The Durata, Riata ST Optim, QuickFlex μ, and Quartet leads should be relatively immune to CCE*. The Durata leads are extremely resistant to longitudinal deformation and probably cause mediastinal displacement rather than differential pulling in response to pectoral movements in vivo. Implantation techniques and lead designs can be used to minimize the risk of CCE*. A bench test for CCE* can be constructed.

This content is not available in your region!

Continue researching here.

Having issues? You can contact us here