Comparison of the pharyngeal pressure provided by two heated, humidified high‐flow nasal cannulae devices in premature infants
Author(s) -
Collins Clare L,
Holberton James R,
König Kai
Publication year - 2013
Publication title -
journal of paediatrics and child health
Language(s) - English
Resource type - Journals
SCImago Journal Rank - 0.631
H-Index - 76
eISSN - 1440-1754
pISSN - 1034-4810
DOI - 10.1111/jpc.12277
Subject(s) - medicine , anesthesia , continuous positive airway pressure , pharynx , airway , surgery , obstructive sleep apnea
Aims This study aims to determine if there is a difference in the pharyngeal pressure, measured as a surrogate for continuous positive distending airway pressure, delivered to premature infants between two commonly used heated, humidified high‐flow nasal cannulae ( HHHFNC ) devices: F isher & P aykel H ealthcare HHHFNC and V apotherm 2000i. Methods Pharyngeal pressure measurements were taken from stable premature infants receiving HHHFNC for respiratory support. Flow rates of 2–8 L/min were studied. Results Nine infants had pharyngeal pressure measurements recorded with both HHHFNC devices at flow rates of 2–8 L/min. There was no difference in pharyngeal pressures recorded between devices at flow rates of 2–6 L/min; measured pressure was linearly associated with flow ( R 2 = 0.9). At flow rates of 7 L/min, V apotherm delivered a mean (standard deviation) pharyngeal pressure of 4.7 (2.2) cmH 2 O compared with 4.23 (2.2) cmH 2 O by the F isher & P aykel device ( P = 0.04). At a flow of 8 L/min, the mean pharyngeal pressure via V apotherm was 4.9 (2.2) cmH 2 O compared with 4.1 (2.3) cmH 2 O with the F isher & P aykel device ( P = 0.05). Conclusions Both HHHFNC delivered similar pharyngeal pressures at flow rates of 2–6 L/min. The pressure limiter valve of the F isher & P aykel device attenuated the pharyngeal pressures at flows of 7 and 8 L/min. V apotherm trended towards higher delivered pharyngeal pressure at flow rates 7 and 8 L/min, but the clinical significance of the difference remains unclear.
Accelerating Research
Robert Robinson Avenue,
Oxford Science Park, Oxford
OX4 4GP, United Kingdom
Address
John Eccles HouseRobert Robinson Avenue,
Oxford Science Park, Oxford
OX4 4GP, United Kingdom