Premium
Robust fluorescent labelling of micropipettes for use in fluorescence microscopy: application to the observation of a mosquito borne parasite infection
Author(s) -
BALABAN AMANDA E.,
NEUMAN KEIR,
SINNIS PHOTINI,
BALABAN ROBERT S.
Publication year - 2018
Publication title -
journal of microscopy
Language(s) - English
Resource type - Journals
SCImago Journal Rank - 0.569
H-Index - 111
eISSN - 1365-2818
pISSN - 0022-2720
DOI - 10.1111/jmi.12610
Subject(s) - pipette , fluorescence , microscopy , fluorescence microscope , microscope , dermis , materials science , biomedical engineering , chemistry , anatomy , biology , optics , pathology , medicine , physics
Summary The ability to monitor micropipette injections with a high‐resolution fluorescent microscope has utility for a variety of applications. Herein, different approaches were tested for creating broad‐band fluorescently labelled glass micropipettes including: UV cured glass glues, baked glass enamel containing fluorescent dyes as well as nanodiamonds attached during pipette formation in the microforge. The most robust and simplest approach was to use labelled baked enamel on the exterior of the pipette. This approach was tested using pipettes designed to mimic a mosquito proboscis for the injection of the malaria parasite, Plasmodium spp., into the dermis of a living mouse ear. The pipette (∼30 micron diameter) was easily detected in the microscopy field of view and tolerated multiple insertions through the skin. This simple inexpensive approach to fluorescently labelling micropipettes will aid in the development of procedures under the fluorescent microscope.