SIDE LOBES AND GRATING LOBES ARTIFACTS IN ULTRASOUND IMAGING
Author(s) -
Paul Y,
Barthez DVM,
Léveillé René,
Peter V,
Scrivani DVM
Publication year - 1997
Publication title -
veterinary radiology and ultrasound
Language(s) - English
Resource type - Journals
SCImago Journal Rank - 0.541
H-Index - 60
eISSN - 1740-8261
pISSN - 1058-8183
DOI - 10.1111/j.1740-8261.1997.tb02104.x
Subject(s) - imaging phantom , transducer , side lobe , optics , grating , acoustics , curvilinear coordinates , artifact (error) , echogenicity , ultrasound , physics , geometry , computer science , mathematics , computer vision , antenna (radio) , telecommunications
Side lobes and grating lobes are both unwanted parts of the ultrasound beam emitted off axis that produce image artifacts due to error in positioning the returning echo. The purpose of this study was to reproduce artifacts associated with side lobes and grating lobes in vitro using different transducer types and recognize these artifacts in vivo. A phantom, composed of a water bath, a metallic wire, and a wooden tongue depressor, was imaged using a linear array, a curved linear array, a vector array, and a sector mechanical transducer. When imaging the metallic wire in a transverse plane, an echogenic artifact was constantly seen on each side of the wire, with a shape and intensity variable with the transducer type. The artifact was curvilinear and concave (linear and curved linear arrays), or curvilinear and convex (vector array and the mechanical transducer). When the tongue depressor was imaged in a longitudinal plane, the artifact was a straight line (linear array), a curved convex line(curved array), a series of convex curvilinear echo (vector array) or a small convex curvilinear echo (mechanical transducer). In vivo situations similar to the phantom experiment were investigated using clinical patients. Artifacts produced in vitro were recognized in vivo when a highly reflective object (urinary bladder wall) was imaged adjacent to an anechoic region (urine). These artifacts corresponded to the principle of secondary ultrasound lobes, and were therefore interpreted as such.

This content is not available in your region!

Continue researching here.

Having issues? You can contact us here