Sensitivity and Precision of Activated Partial Thromboplastin Time (APTT) Methods
Author(s) -
Refsum Nina,
Collen D.,
Godal H. C.,
Janson T.,
Mannucci P. M.,
Nilsson Inga Marie,
GilhuusMoe C. C.
Publication year - 1978
Publication title -
scandinavian journal of haematology
Language(s) - English
Resource type - Journals
SCImago Journal Rank - 0.904
H-Index - 84
eISSN - 1600-0609
pISSN - 0036-553X
DOI - 10.1111/j.1600-0609.1978.tb01559.x
Subject(s) - partial thromboplastin time , prothrombin time , medicine , coagulation , coagulation testing , mathematics , anesthesia , chromatography , chemistry , surgery
The Activated Partial Thromboplastin Time (APTT) test, Cephotest, was compared to other APTT methods in current use in 4 specialized coagulation laboratories. In 3 of 4 laboratories, the sensitivity of Cephotest was superior (P < 0.001) to that of the local APTT method. There was no statistically significant difference between the APTT methods with regard to precision of repetitive testing. In each laboratory, the normal range of Cephotest was estimated on freshly collected plasma samples from healthy subjects. A mean value between 28.8 and 35.8 s, with a standard deviation of 1.1–3.3 s, was obtained. It is concluded that the composition of the APTT method is of importance for the sensitivity of this test, but does not influence the precision of repetitive testing to a significant degree. The use of a standardized reagent facilitates comparison of the results obtained with the APTT method from one laboratory to another.