The effect of working parameters on root substance removal using a piezoelectric ultrasonic sealer in vitro
Author(s) -
Flemmig T. F.,
Petersilka G. J.,
Mehl A.,
Hickel R.,
Klaiber B.
Publication year - 1998
Publication title -
journal of clinical periodontology
Language(s) - English
Resource type - Journals
SCImago Journal Rank - 3.456
H-Index - 151
eISSN - 1600-051X
pISSN - 0303-6979
DOI - 10.1111/j.1600-051x.1998.tb02422.x
Subject(s) - ultrasonic sensor , piezoelectricity , materials science , volume (thermodynamics) , biomedical engineering , instrumentation (computer programming) , ultrasound , composite material , dentistry , acoustics , medicine , computer science , physics , quantum mechanics , operating system
. This study assessed defect depth and volume resulting from root instrumentation using a piezoelectric ultrasonic sealer with a slim scaling tip in vitro. Combinations of the following working parameters were analyzed: lateral forces of 0.5 N, 1 N, and 2 N; tip angulations of O°. 45°. and 90°; power settings of low. medium and high: and instrumentation time of 10 s. 20 s. 40 s. and 80 s. Defects were quantified using a 3D optical laser scanner. Overall, lateral force had the greatest influence on defect volume compared to instrument power setting and tip angulation (β‐weights 0.49±0.04, 0.25±0.D4, and 0.14±0.04, respectively). The effects on defect depth were highest for tip angulation followed by lateral force and instrument power setting (β‐weights 0.48±0.04. 0.34±0.04, and 0.25±0.04, respectively). Interestingly, al all power settings, the highest defect volume and depth by far were found after combining 45° tip angulation with 2 N of lateral force. The efficacy of the assessed piezoelectric ultrasonic sealer may be adapted to the various clinical needs by adjusting the lateral force. tip angulation, and power setting. To prevent severe root damage it is crucial to use the assessed sealer at a tip angulation of close to 0°.