Validation of Powder X‐Ray Diffraction Following EN ISO/IEC 17025
Author(s) -
Eckardt Regina,
Krupicka Erik,
Hofmeister Wolfgang
Publication year - 2012
Publication title -
journal of forensic sciences
Language(s) - English
Resource type - Journals
SCImago Journal Rank - 0.715
H-Index - 96
eISSN - 1556-4029
pISSN - 0022-1198
DOI - 10.1111/j.1556-4029.2011.02025.x
Subject(s) - powder diffraction , materials science , robustness (evolution) , analytical chemistry (journal) , chromatography , chemistry , crystallography , biochemistry , gene
Powder X‐ray diffraction (PXRD) is used widely in forensic science laboratories with the main focus of qualitative phase identification. Little is found in literature referring to the topic of validation of PXRD in the field of forensic sciences. According to EN ISO/IEC 17025, the method has to be tested for several parameters. Trueness, specificity, and selectivity of PXRD were tested using certified reference materials or a combination thereof. All three tested parameters showed the secure performance of the method. Sample preparation errors were simulated to evaluate the robustness of the method. These errors were either easily detected by the operator or nonsignificant for phase identification. In case of the detection limit, a statistical evaluation of the signal‐to‐noise ratio showed that a peak criterion of three sigma is inadequate and recommendations for a more realistic peak criterion are given. Finally, the results of an international proficiency test showed the secure performance of PXRD.