Advanced ICD Troubleshooting: Part I
Author(s) -
SWERDLOW CHARLES D.,
FRIEDMAN PAUL A.
Publication year - 2005
Publication title -
pacing and clinical electrophysiology
Language(s) - English
Resource type - Journals
SCImago Journal Rank - 0.686
H-Index - 101
eISSN - 1540-8159
pISSN - 0147-8389
DOI - 10.1111/j.1540-8159.2005.00275.x
Subject(s) - medicine , troubleshooting , citation , library science , associate editor , computer science , operating system
Failure to Deliver Therapy or Delay of Therapy Absent or delayed therapy may be caused by programmed values (including human error), ICD system performance, or a combination of the two. The most common causes are ICD inactivation, VT slower than the programmed detection interval, SVT-VT discriminators, undersensing, and intradevice software interactions. Pacemaker-ICD interactions, once important, are now rare due to incorporation of dualand triple-chamber pacemakers in ICDs.