Author(s) -
CAMOUS JEAN PIERRE,
RAYBAUD FLORENCE,
JOURDAN JACQUES
Publication year - 1999
Publication title -
pacing and clinical electrophysiology
Language(s) - English
Resource type - Journals
SCImago Journal Rank - 0.686
H-Index - 101
eISSN - 1540-8159
pISSN - 0147-8389
DOI - 10.1111/j.1540-8159.1999.tb00395.x
Subject(s) - medicine , lead (geology) , abdomen , cardiology , surgery , geology , geomorphology
A proximal pseudofracture of a Medtronic 6936 defibrillating lead adjacent to the ICD pulse generator was found on X ray. At this site, the multifilar metallic connector joins the stimulating lead before coiling around it. Conventional X ray of the abdomen does not visualize this connection. During ICD replacement, lead impedance should be routinely measured.