z-logo
Premium
Limitations of the Algorithm to Reconfirm Ventricular Fibrillation in a Third‐Generation Implantahle Cardioverter Defibrillator
Author(s) -
MATTKE SÖREN,
KRAMM BERTHOLD
Publication year - 1997
Publication title -
pacing and clinical electrophysiology
Language(s) - English
Resource type - Journals
SCImago Journal Rank - 0.686
H-Index - 101
eISSN - 1540-8159
pISSN - 0147-8389
DOI - 10.1111/j.1540-8159.1997.tb05507.x
Subject(s) - medicine , ventricular fibrillation , cardiology , implantable cardioverter defibrillator , bradycardia , shock (circulatory) , algorithm , interval (graph theory) , fibrillation , atrial fibrillation , heart rate , combinatorics , mathematics , computer science , blood pressure
We report on inadequate shocks delivered by an ICD for VT converting spontaneously to AF with slow ventricular response. This represents a correct behavior of the reconfirmation algorithm, as the detection of intervals longer than the fibrillation detection interval and shorter than the bradycardia escape interval is required to inhibit the shock.

This content is not available in your region!