Distortion of Intracardiac Electrograms Following Defibrillator Shocks for Atrial Tachyarrhythmias
Author(s) -
KHALIGHI KOROUSH,
FLORIN TODD J.,
PETERS ROBERT W.,
SHOROFSKY STEPHEN R.,
GOLD MICHAEL R.
Publication year - 1997
Publication title -
pacing and clinical electrophysiology
Language(s) - English
Resource type - Journals
SCImago Journal Rank - 0.686
H-Index - 101
eISSN - 1540-8159
pISSN - 0147-8389
DOI - 10.1111/j.1540-8159.1997.tb03539.x
Subject(s) - medicine , intracardiac injection , cardiology , ventricular tachycardia , tachycardia , distortion (music) , electrical engineering , amplifier , cmos , engineering
In three patients with a defibrillator system consisting of a Ventak P2 pulse generator and an Endotak C transvenous lead, we observed distortion of introcardiac electrograms following defibrillator shocks for atrial arrhythmias. There was a transient marked widening of the intracardiac ventricular complexes resembling ventricular tachycardia. This phenomenon should be recognized when evaluating arrhythmic episodes.