Impedance of Pacemaker Leads : Correlation of Different Methods
Author(s) -
SIEGMUND JOACHIM B.,
WILSON JOHN H.,
LATTNER STEPHEN E.,
GRANNEMAN KIMBERLY,
JOHNSON ROBERT
Publication year - 1996
Publication title -
pacing and clinical electrophysiology
Language(s) - English
Resource type - Journals
SCImago Journal Rank - 0.686
H-Index - 101
eISSN - 1540-8159
pISSN - 0147-8389
DOI - 10.1111/j.1540-8159.1996.tb04795.x
Subject(s) - medicine , impulse (physics) , electrical impedance , correlation , focused impedance measurement , cardiology , biomedical engineering , electrical engineering , mathematics , physics , geometry , quantum mechanics , engineering
This study compares three different pacing system analyzers, which measure impedance at different points during an impulse, with measurements telemetered from implanted pacemakers from four different manufacturers. Measurements were obtained at the time of implantation in a group of 103 patients. The measurements obtained by these different methods differ significantly from each other; the later during the impulse the impedance is measured, the higher in general its value. Interpretation of impedance values should take into account the technique of measurement.