Large Capacitor Defibrillation Waveform Reduces Peak Voltages Without Increasing Energies
Author(s) -
HAHN STEPHEN J.,
HEIL JOHN E.,
LANG DOUGLAS J.
Publication year - 1995
Publication title -
pacing and clinical electrophysiology
Language(s) - English
Resource type - Journals
SCImago Journal Rank - 0.686
H-Index - 101
eISSN - 1540-8159
pISSN - 0147-8389
DOI - 10.1111/j.1540-8159.1995.tb02505.x
Subject(s) - medicine , defibrillation threshold , defibrillation , capacitor , waveform , capacitance , voltage , cardiology , anesthesia , electrode , electrical engineering , physics , quantum mechanics , engineering
This study tested the hypothesis that increasing capacitance would allow a reduction in ICD size without reducing the deliverable energy. For example, the volume of a single 450 μF capacitor (390 V peak) is 1/3 less than that of two 250 μF capacitors (780 V), but it can store equivalent amounts of energy.Methods: Endocardial defibrillation electrodes (3.4 cm) were positioned in the RV apex and at the RA/ SVC junction in six mixed‐breed, isoflurane anesthetized pigs (41 ± 3 kg). Three 17‐cm ribbon wires were positioned subcutaneously on the left lateral chest (SQArray). Two CPI VENTAK ECDs were equipped to deliver 60/40 biphasic waveforms using either 125 μF (STD) or 500 μF (LD) of capacitance. A 15 shock up/down protocol was used to determine the 50% probability of success levels for each waveform in each animal. Shocks were delivered from RV(‐)→SVC + SQArray(+) in random order. Results were compared using paired Student'sf‐tests and are reported as mean ± SE. Results: The 500 μF long duration waveform reduced peak voltage 41% (374 ± 18 V [STD] vs 219 ± 14 V (LD], P < 0.001) and reduced peak current 38% (11.0 ± 1.1 A [STD] vs 6.8 ± 0.6 A [LD], P < 0.001) but did not significantly change the delivered energy(12.4 ± 1.3 J [STD] vs 13.4 ± 1.0 J [LD]). Durations increased from 10.0 ± 0.2 to 17.6 ± 0.5 msec (P < 0.001).Conclusions: Defibrillation with a 500 μF, long duration biphasic defibrillation waveform received similar amounts of energy but significantly reduced peak voltage and current compared to a waveform produced from 125 μF. A single large capacitor could be used to reduce the physical size of an ICD compared to the standard two capacitor system.

This content is not available in your region!

Continue researching here.

Having issues? You can contact us here