First Derivative of Right Ventricular Pressure, dP/dt, as a Sensor for a Rate Adaptive VVI Pacemaker: Initial Experience
Author(s) -
OVSYSHCHER ILYA,
GUETTA VICTOR,
BONDY CHAVA,
PORATH AVI
Publication year - 1992
Publication title -
pacing and clinical electrophysiology
Language(s) - English
Resource type - Journals
SCImago Journal Rank - 0.686
H-Index - 101
eISSN - 1540-8159
pISSN - 0147-8389
DOI - 10.1111/j.1540-8159.1992.tb03065.x
Subject(s) - medicine , ventricular pacing , lead (geology) , cardiology , heart rate , ventricular pressure , anesthesia , artificial cardiac pacemaker , blood pressure , heart failure , geomorphology , geology
Ten patients underwent implantation of a rate adaptive ventricular pacing system with a new pulse generator and lead. The unipolar lead has a steroid eluting tip and a pressure sensor. The first derivative of the signal from this sensor, dP/dt, is determined and the pacemaker rate is varied in response to changes in the right ventricular dP/dt MAx . During implantation, dP/dt values were in the range of 180–720 mm Hg/sec, The autothreshold for pacing at 2.5 V remained unchanged 1 month after implantation (0.065 ± 0.045 msec, range 0.05–2.00 msec) and only slightly increased after 3 months (0.075 ± 0.045 msec, range 0.05–2.00 msec). A significant correlation existed between the dP/dt measured during implantation and the right ventricular pressure measured by telemetry at follow‐up visits (r = 0.93, P = 0.0001). Initial pacemaker programming was performed on the second day after implantation following a short walk and was adjusted subsequent to follow‐up visits according to the patient's subjective assessment and in accordance with the results of exercise tests and Holter monitoring. Exercise and Holter tests did not significantly change initial programming. There was a significant correlation between right ventricular systolic pressure and the rate response setting (r = ‐ 0.66, P < 0.05). During dP/dt pacing, all patients felt well, and eight of these reported an improvement compared to non rate adaptive pacing. The heart rate response to effort and recovery was appropriate. It was concluded that: (1) right ventricular dP/dt is a suitable parameter for controlling the pacing rate; (2) appropriate programming of the dP/dt pacemaker results in a suitable heart rate response to exercise and recovery.

This content is not available in your region!

Continue researching here.

Having issues? You can contact us here