z-logo
Premium
Double Sensing by Atrial Automatic Tacbycardia‐Terminating Pulse Generator
Author(s) -
BAROLD S. SERGE,
FALKOFF MICHAEL D.,
ONG LING S.,
HEINLE ROBERT A.
Publication year - 1987
Publication title -
pacing and clinical electrophysiology
Language(s) - English
Resource type - Journals
SCImago Journal Rank - 0.686
H-Index - 101
eISSN - 1540-8159
pISSN - 0147-8389
DOI - 10.1111/j.1540-8159.1987.tb05924.x
Subject(s) - medicine , supraventricular tachycardia , cardiology , qrs complex , tachycardia , sinus rhythm , normal sinus rhythm , pulse (music) , pulse generator , normalization property , atrial tachycardia , ventricular tachycardia , electrocardiography , sinus tachycardia , anesthesia , atrial fibrillation , electrical engineering , voltage , computer science , catheter ablation , programming language , engineering
This report describes two cases of double sensing of the P and QRS signals by an implanted atrial bipolar automatic burst tachycardia‐terminating puise generator (Inlermedics CyherTach 60) used for the treatment of supraventricular tachycardia. Double sensing occurred during normai sinus rhythm at a rate slower than the tachycardia detection criterion and caused inappropriate delivery of burst stimulation. These observations underscore the importance of the far‐field QRS signal in the detection of supraventricular tachycardia by automatic puise generators sensing the atrial eiectrogram.