Failure to Sense P Waves During DDD Pacing
Author(s) -
MECHELEN ROB,
'T HART CONNIE,
BOER HANS
Publication year - 1986
Publication title -
pacing and clinical electrophysiology
Language(s) - English
Resource type - Journals
SCImago Journal Rank - 0.686
H-Index - 101
eISSN - 1540-8159
pISSN - 0147-8389
DOI - 10.1111/j.1540-8159.1986.tb06605.x
Subject(s) - medicine , cardiology , t wave , electrocardiography
ECG tracings of three patients in whom AV universal (DDD) pacemakers were implanted intermittently demonstrated dropped P waves. In one patient, true atrial undersensing was present; in the others, sensing of the atrial electrode was appropriate, but sensing of sinus P waves was intermittently blocked by normal pacemaker operation. In this report we discuss the electrocardiographic diagnosis of atrial undersensing in order to avoid unnecessary reinterventions.