Evaluation of immunohematologic routine methods using the new erythrocyte‐magnetized technology on the QWALYS 2 system
Author(s) -
Schoenfeld Helge,
Bulling Katia,
Heymann Christian von,
Neuner Bruno,
Kalus Ulrich,
Kiesewetter Holger,
Pruss Axel
Publication year - 2009
Publication title -
transfusion
Language(s) - English
Resource type - Journals
SCImago Journal Rank - 1.045
H-Index - 132
eISSN - 1537-2995
pISSN - 0041-1132
DOI - 10.1111/j.1537-2995.2009.02136.x
Subject(s) - abo blood group system , concordance , antibody , rh blood group system , centrifugation , typing , red blood cell , chromatography , chemistry , immunology , medicine , biology , genetics
BACKGROUND: QWALYS 2 is a fully automated system for ABO/D grouping, Rh phenotyping, K typing, and antibody screening (ABS). Its new erythrocyte‐magnetized technology (EMT) is based on the use of magnetic nanoparticles and avoids centrifugation and washing steps. STUDY DESIGN AND METHODS: Overall 499 blood samples were tested with our routine blood bank methods for ABO/D grouping, 313 samples for Rh phenotyping and K typing (microtiter plates; Olympus PK 7200), and 478 samples for ABS (gel centrifugation technique, DiaMed). All samples were tested in parallel with the EMT. RESULTS: In 496 of 499 samples (99.4%), a complete concordance between the observed (QWALYS 2) and the expected results for ABO/D grouping was found. One sample with a weak A in an AB blood group and 2 samples with a weak D were not detected by the QWALYS system. Rh phenotyping and K tests revealed a 100% concordance. In the two ABS techniques, 427 samples were negative in both and 15 samples showed the same antibody specificity in both. Three immunoglobulin M antibodies were as expected negative in EMT and positive by DiaMed. In 32 cases (6.7%), false‐positive reactions were observed by EMT due to 22 unspecific reactions (4.6%) and 10 lipemic or fibrinic plasmas (2.1%). One autoantibody was found by EMT only. CONCLUSION: The EMT is reliably suited to ABO/D grouping, Rh phenotyping, and K testing and is suitable to detect immunoglobulin G red blood cell alloantibodies as well. The rate of false‐positive reactions in ABS due to lipemic and fibrinic samples needs to be reduced.