z-logo
Premium
The Vienna Implantable Centrifugal Blood Pump
Author(s) -
SCHIMA H.,
TRUBEL W.,
WIESELTHALERXY G.,
SCHMIDT C.,
MÜLLER M. R.,
SIEGL H.,
LOSERT U.,
WOLNER E.
Publication year - 1994
Publication title -
artificial organs
Language(s) - English
Resource type - Journals
SCImago Journal Rank - 0.684
H-Index - 76
eISSN - 1525-1594
pISSN - 0160-564X
DOI - 10.1111/j.1525-1594.1994.tb03367.x
Subject(s) - impeller , centrifugal pump , biomedical engineering , hemolysis , thrombus , computer science , materials science , medicine , surgery , mechanical engineering , engineering
Because of the inherent disadvantages of membrane pumps, rotary pumps have been increasingly investigated in recent years. As a result of improving biocompatibility, extended assistance with implantable devices is of special interest. Questions arise concerning shear stress, blood traumatization, design of seals, and specific control conditions. In their development of an implantable impeller pump, the Vienna group studied the minimization of hemolysis and thrombus formation by means of numerical simulation, visualization, and in vitro blood evaluation. The latter was revealed to be the most powerful tool for pump evaluation. With optimization of geometry, a hemolysis of in vitro: IH = 0. 008; MIH = 0. 58; and in vivo: 2. 1 to 3 mg% plasma–free hemoglobin could be obtained. For proper control and physiological adaptation, a controller based on a nonlinear and a fuzzy strategy was developed. Furthermore, a method for evaluation of the contractility of the assisted heart during nonpulsatile support was tested by computer simulation. This paper summarizes the evaluation methods used and provide an overview of the results of pump and controller design.

This content is not available in your region!

Continue researching here.

Having issues? You can contact us here