Continuous Hematocrit Monitoring Method in an Extracorporeal Circulation System and Its Application for Automatic Control of Blood Volume during Artificial Kidney Treatment
Author(s) -
Ishihara Toshikazu,
Igarashi Isemi,
Kitano Tomoyuki,
Shinzato Takahiro,
Maeda Kenji
Publication year - 1993
Publication title -
artificial organs
Language(s) - English
Resource type - Journals
SCImago Journal Rank - 0.684
H-Index - 76
eISSN - 1525-1594
pISSN - 0160-564X
DOI - 10.1111/j.1525-1594.1993.tb00619.x
Subject(s) - extracorporeal circulation , hematocrit , artificial kidney , blood volume , medicine , volume (thermodynamics) , blood circulation , extracorporeal , biomedical engineering , kidney , cardiology , physics , quantum mechanics , traditional medicine
A two‐frequency electrical conductivity method for measuring hematocrit has been developed for use in the continuous monitoring of circulating blood volume during artificial kidney treatment. Using this method, an algorithm for programmed control of blood volume also has been developed. This hematocrit measurement method is based on the β‐dispersion of blood conductivity, and the values measured during hemodialysis are in good agreement with the laboratory hematocrit values obtained by a Coulter counter. Programmed control of blood volume was used 29 times on 11 patients undergoing long‐term dialysis. In each case, the blood volume was maintained within 2.5% of the desired level, and stable control was achieved even with disturbances caused by eating, change in body position, and so on. These methods can be used to study the relationship between blood volume and dialysis‐induced hypotension during artificial kidney treatment and to achieve adequate ultrafiltration without clinical adverse effects.

This content is not available in your region!

Continue researching here.

Having issues? You can contact us here