z-logo
Premium
Accuracy of Physical Examination and Intra‐Access Pressure in the Detection of Stenosis in Hemodialysis Arteriovenous Fistula
Author(s) -
Campos Rodrigo Peixoto,
Chula Domingos Candiota,
Perreto Sonia,
Riella Miguel Carlos,
Do Nascimento Marcelo Mazza
Publication year - 2008
Publication title -
seminars in dialysis
Language(s) - English
Resource type - Journals
SCImago Journal Rank - 0.899
H-Index - 78
eISSN - 1525-139X
pISSN - 0894-0959
DOI - 10.1111/j.1525-139x.2007.00419.x
Subject(s) - medicine , hemodialysis access , arteriovenous fistula , hemodialysis , stenosis , vascular access , physical examination , radiology , cardiology
Both physical examination (PE) and intra‐access pressure (IAP) measurements have been used in the identification of stenosis in an arteriovenous access. The aim of this study was to evaluate the accuracy of PE and IAP in the diagnosis of arteriovenous fistula (AVF) stenosis. A total of 84 patients were enrolled in the study (54% men, mean age of 50.7 ± 12.7 years and mean AVF patency of 24.9 ± 7.8 months, 52% radiocephalic). Abnormalities of pulse and thrill were used as the diagnostic tools for the detection of stenosis using the physical examination. For IAP, stenosis was suspected when the ratio between IAP at the arterial puncture site and the mean blood pressure was <0.13 or >0.43. The diagnosis of stenosis was confirmed by Doppler ultrasound (DU). Sensitivity (S), specificity (SP), positive predictive value (PPV), negative predictive value (PNV), and accuracy were calculated for the two early detection tests. According to DU, 50 (59%) AVF were considered positive for the presence of stenosis. Fifty‐six (66%) AVF were considered positive for the presence of stenosis by PE and 34 (40%) by IAP. S, SP, PPV, and NPV for PE and IAP were 96%, 76%, 86%, and 93% and 60%, 88%, 88%, and 60%, respectively. The accuracy for PE and IAP was 88% and 71%, respectively. PE proved to be an accurate method for the diagnosis of stenosis and should be part of all surveillance protocols of stenosis detection in AVF.

This content is not available in your region!

Continue researching here.

Having issues? You can contact us here