Rapid Ward Assessment of Renal Function by Reflectance Photometric Determination of Plasma Creatinine
Author(s) -
GIBB I.,
CROSS P.,
PRATT D.,
HALL R. R.
Publication year - 1985
Publication title -
british journal of urology
Language(s) - English
Resource type - Journals
SCImago Journal Rank - 1.773
H-Index - 148
eISSN - 1464-410X
pISSN - 0007-1331
DOI - 10.1111/j.1464-410x.1985.tb05857.x
Subject(s) - renal function , creatinine , reflectivity , medicine , urology , optics , physics
Summary— The Ames Seralyzer reflectance photometer and disposable reagent strips for the determination of plasma creatinine were evaluated in the laboratory and on a urology ward over a 3‐month period. Principles of the instrument used are described, as are details of the evaluation procedures. Reproducibility of results under optimum conditions gave coefficients of variation < 7%. Under routine conditions of use, medical staff and laboratory analysts achieved high correlation with the routine laboratory method of analysis (r = 0.986, N = 108: r = 0.996, N = 102 respectively). Potential sources of inaccuracy were identified. Seralyzer determination of plasma creatinine can be rapidly, simply and reliably performed by medical staff on a urology ward, thereby facilitating assessment of renal function governed by clinical requirement rather than by laboratory convenience.