In‐vivo Determination of Error in the Measurement of Urethral Pressure by the Method of Brown and Wickham
Author(s) -
BROWN MALCOLM
Publication year - 1975
Publication title -
british journal of urology
Language(s) - English
Resource type - Journals
SCImago Journal Rank - 1.773
H-Index - 148
eISSN - 1464-410X
pISSN - 0007-1331
DOI - 10.1111/j.1464-410x.1975.tb04004.x
Subject(s) - in vivo , mathematics , biology , microbiology and biotechnology
Summary A method of determining the error caused by the infusion of water into the urethra during urethral wall pressure and urethral pressure profile measurements is described and discussed. The test is conducted in vivo and the measured pressure is related to the actual urethral wall pressure. The “intrinsic error” is found to be approximately 1·0 cm H 2 O for every ml/min of infusion rate. Incidental findings concerned with changes in urethral resistance are also reported.