RIGHT ATRIAL CATHETERS: RELIABLE, IMMEDIATE AND DURABLE VASCULAR ACCESS FOR HAEMODIALYSIS AND PLASMA EXCHANGE
Author(s) -
Francis D. M. A.,
Goodman D. J.,
Millar R. J.,
Walker R. G.,
Becker G. J.
Publication year - 1993
Publication title -
australian and new zealand journal of surgery
Language(s) - English
Resource type - Journals
SCImago Journal Rank - 0.111
H-Index - 51
eISSN - 1445-2197
pISSN - 0004-8682
DOI - 10.1111/j.1445-2197.1993.tb00384.x
Subject(s) - medicine , catheter , external jugular vein , surgery , vascular access , subclavian vein , lumen (anatomy) , silastic , cardiology , hemodialysis
This paper prospectively evaluates 33 dual lumen, right atrial catheters inserted into either an external or internal jugular vein by open operation in 29 patients, of whom 15 required haemodialysis and 14 required temporary plasma exchange. The median (range) catheter survival in the haemnodialysis and plasma exchange groups was 108 days (7–334 days) and 61 days (10–116 days), respectively. Life table analysis demonstrated that overall catheter survival was 58% at 200 days. The main causes of catheter failure were infection (four cases), poor flow (three cases) and accidental removal (one case). Another nine catheters were removed electively because of maturation of alternative methods of vascular access (five cases). completion of plasma exchange treatment (three cases), or successful renal transplantation (one case). Long‐term silastic catheters, inserted into the right atrium via a jugular vein. have distinct advantages over temporary subclavian vein catheters and external arteriovenous (AV) shunts; this form of access is the method of choice for hamodialysis and plasma exchange patients who require immediate and short‐ to medium‐term vascular access.