Fatal pneumoperitoneum caused by nasopharyngeal oxygen delivery after transoesophageal echocardiography for cardiac surgery
Author(s) -
MOURISSOUX G.,
SCHLUMBERGER S.,
DE LENTDECKER P.,
FISCHLER M.
Publication year - 2009
Publication title -
acta anaesthesiologica scandinavica
Language(s) - English
Resource type - Journals
SCImago Journal Rank - 0.738
H-Index - 107
eISSN - 1399-6576
pISSN - 0001-5172
DOI - 10.1111/j.1399-6576.2009.02056.x
Subject(s) - medicine , pneumoperitoneum , autopsy , cannula , surgery , laparoscopy
We report a case of fatal post‐operative pneumoperitoneum in a patient who had undergone urgent mitral valve surgery. In the absence of a proven cause of the pneumoperitoneum (refusal by the family of an autopsy), we can only propose a hypothesis for its origin. The most probable one is that forceful or sustained retrograde flexion of the transoesophageal echocardiographic probe created a lower oesophagus or gastric rupture and that oxygen flow administered by the nasal cannula went straight to the abdominal cavity, leading to tension pneumoperitoneum.