RELIABILITY OF HALOTHANE VAPORIZERS
Author(s) -
Adner Marc,
Hallén Börje
Publication year - 1965
Publication title -
acta anaesthesiologica scandinavica
Language(s) - English
Resource type - Journals
SCImago Journal Rank - 0.738
H-Index - 107
eISSN - 1399-6576
pISSN - 0001-5172
DOI - 10.1111/j.1399-6576.1965.tb00510.x
Subject(s) - medicine , halothane , gynecology , microbiology and biotechnology , anesthesia , biology
SUMMARY The function of a number of vaporizers, new and used, were tested at different flows and different dial settings. The vapour concentrations produced were measured by means of a gas‐chromatographic technique. The results of the tests show that the preset concentrations are not always identical with the concentrations produced. ZUSAMMENFASSUNG Eine Anzahl von neuen und gebrauchten Verdampfern wurde bei verschiedenen Durchströmungsmengen und verschiedenen Einstellungen hinsichtlich ihrer Funktion getestet. Die erzielten Dampfkonzentrationen wurden mit Hilfe der Gaschromatographie gemessen. Die Ergebnisse dieser Untersuchungen zeigen, dass die eingestellten Kon‐zentrationswerte nicht immer mit den gelieferten Konzentrationen identisch sind.