Subject(s) - medicine , anesthesia , cyclopropane , oxygen , ring (chemistry) , chemistry , organic chemistry
SUMMARY Oxygen concentrations were measured in the gas mixtures of cyclopropane and oxygen delivered to patients during a short operation (35 minutes). It has been demonstrated that the amount of oxygen actually administered is considerably lower than it appears on the flowmeters of the anesthesia apparatus. ZUSAMMENFASSUNG Die Sauerstoffkonzentrationen in Gasgemischen von Cyclopropan und Sauerstoff, welche man Patienten während kurzer Eingriffe von etwa 35 min Dauer verabreichte, wurden gemessen; Es konnte gezeigt werden, dass die tatsächliche Sauerstoffkonzentration beträchtlich unterhalb jener Werte liegt, die von den Flowmetern des Narkoseapparates angezeigt wurden.