Premium
Parameters affecting imaging of the horseradish‐peroxidase‐diaminobenzidine reaction product in the confocal scanning laser microscope
Author(s) -
Deitch Jeffrey S.,
Smith Karen L.,
Swann John W.,
Turner James N.
Publication year - 1990
Publication title -
journal of microscopy
Language(s) - English
Resource type - Journals
SCImago Journal Rank - 0.569
H-Index - 111
eISSN - 1365-2818
pISSN - 0022-2720
DOI - 10.1111/j.1365-2818.1990.tb03062.x
Subject(s) - horseradish peroxidase , confocal , microscope , materials science , laser , confocal laser scanning microscopy , confocal laser scanning microscope , chemistry , optics , biomedical engineering , medicine , biochemistry , physics , enzyme
SUMMARY Neurons intracellularly filled with biocytin and labelled with nickel‐intensified diaminobenzidine (DAB/Ni) can be imaged on the confocal scanning laser microscope in order to obtain three‐dimensional and optical section images of neurons. Intensification of the DAB reaction product with nickel was found to be crucial for obtaining a workable signal level. On the other hand, the high absorption of light by the reaction product severely attenuated the detection of structures lying directly underneath, and the intensity of the unattenuated laser used for imaging faded or damaged the DAB/Ni reaction product. We have determined that reduction of the laser intensity combined with the use of proper objective lenses and non‐laser‐based imaging for preliminary adjustments of the specimen all work to reduce or eliminate damage, and also improve the image. These items must be kept in mind when imaging and analysing DAB‐labelled structures in the laser‐based confocal microscope.