Premium
A randomised trial comparing the CEL‐100 videolaryngoscope TM with the Macintosh laryngoscope blade for insertion of double‐lumen tubes
Author(s) -
Lin W.,
Li H.,
Liu W.,
Cao L.,
Tan H.,
Zhong Z.
Publication year - 2012
Publication title -
anaesthesia
Language(s) - English
Resource type - Journals
SCImago Journal Rank - 1.839
H-Index - 117
eISSN - 1365-2044
pISSN - 0003-2409
DOI - 10.1111/j.1365-2044.2012.07137.x
Subject(s) - medicine , intubation , lumen (anatomy) , surgery , tracheal intubation , anesthesia , nuclear medicine
Summary We performed a randomised trial comparing the CEL‐100 videolaryngoscope TM with the Macintosh laryngoscope blade in 170 patients undergoing double‐lumen tube placement for thoracic surgery. Compared with the Macintosh laryngoscope blade, use of the CEL‐100 resulted in significantly more patients with a Cormack and Lehane Grade‐1 laryngeal view (90.4% vs 61.0%, p < 0.001), a higher rate of successful intubation on the first attempt (92.8% vs 79.3%, p = 0.012), a lower median (IQR [range]) intubation difficulty score (0 (0–0 [0–60]) vs 15 (0–30 [0–80]), p < 0.001), a higher incidence of correct positioning of the tube (90.3% vs 79.2%, p = 0.041) and significantly fewer patients requiring external laryngeal pressure (19.3% vs 32.9%, p = 0.046). Median (IQR [range]) time to successful intubation was 45 (38–55 [22–132]) s with the CEL‐100 compared with 51 (40–61 [30–160] s using the Macintosh laryngoscope blade. We conclude that the CEL‐100 videolaryngoscope is superior to the Macintosh laryngoscope blade for double‐lumen tube insertion.