The estimation of inspired isoflurane concentration in a low‐flow system
Author(s) -
GOWRIEMOHAN S.,
CHAKRABARTI M. K.,
LOCKWOOD G. G.
Publication year - 1995
Publication title -
anaesthesia
Language(s) - English
Resource type - Journals
SCImago Journal Rank - 1.839
H-Index - 117
eISSN - 1365-2044
pISSN - 0003-2409
DOI - 10.1111/j.1365-2044.1995.tb06099.x
Subject(s) - isoflurane , vaporizer , medicine , fresh gas flow , anesthesia , ventilation (architecture) , mechanical ventilation , sevoflurane , mechanical engineering , pathology , engineering
Summary We have examined the predictability of inspired isoflurane concentration during low‐flow anaesthesia using a to‐and‐fro breathing system. Twenty one adult patients requiring mechanical ventilation of the lungs during surgery took part in this study. Using a fresh gas flow of 2 l.min ‐1 , the ratio of inspired isoflurane concentration to isoflurane vaporizer setting was found to be approximately 4/5th after 5 min of anaesthesia. The ratio was maintained throughout the procedure, except for a few minutes following each change in vaporizer setting.