The output of an‘empty’ Tec 3 vaporizer
Author(s) -
GRAHAM S. G.,
SZAFRANSKI J.,
BLAND D.,
LANGTON J. A.
Publication year - 1995
Publication title -
anaesthesia
Language(s) - English
Resource type - Journals
SCImago Journal Rank - 1.839
H-Index - 117
eISSN - 1365-2044
pISSN - 0003-2409
DOI - 10.1111/j.1365-2044.1995.tb06000.x
Subject(s) - medicine , vaporizer , tec , astronomy , ionosphere , physics , pathology
Summary The output of 30 Tec 3 vaporizers (halothane, enflurane and isoflurane) was studied, starting at the point where no liquid was visible in the content window. At 6l.min ‐1 and 1% v/v initial output, consistent delivery was on average maintained for in excess of 90 (range 55–120) min. Thereafter, output declined rapidly. At the flows and concentrations studied there is a significant reserve in the vaporizing chamber, but it is likely that when higher flows and concentrations are used this reserve may be substantially reduced.