Whole blood electrolyte assay using ChemPro 500
Author(s) -
THOMPSON J. M.,
EMMETT C.,
SMITH S. C. H.,
CRAMB R.,
HUTTON P.
Publication year - 1991
Publication title -
anaesthesia
Language(s) - English
Resource type - Journals
SCImago Journal Rank - 1.839
H-Index - 117
eISSN - 1365-2044
pISSN - 0003-2409
DOI - 10.1111/j.1365-2044.1991.tb09722.x
Subject(s) - medicine , replicate , whole blood , analyser , intensive care unit , electrolyte , confidence interval , quality assurance , chromatography , emergency medicine , surgery , pathology , statistics , chemistry , mathematics , electrode , external quality assessment
Summary The ChemPro 500‘near‐the‐patient’analyser, with ChemPro‘Ion Profile’sensor cards, was evalutated for the assay of pH, Ca 2+ , K + and Na + in whole blood samples from patients in the intensive care unit or during surgery for heart or major blood vessel disease, or for liver transplantation. Imprecisions estimated from replicate whole blood measurements were much greater for all four ions than even the least stringent of the generally accepted analytical goals, and much greater than those estimated using quality assurance materials. Comparisons of assayed values with those obtained using standard laboratory instruments showed significant constant and proportional biases. The performance of the ChemPro 500 with the Ion Profile cards gave us no confidence in recommending their use to anaesthetists and intensivists.