Premium
The Afya anaesthesia system
Author(s) -
SWAI E.A.,
BROOKS A.M.,
MECKLENBURGH J.S.
Publication year - 1985
Publication title -
anaesthesia
Language(s) - English
Resource type - Journals
SCImago Journal Rank - 1.839
H-Index - 117
eISSN - 1365-2044
pISSN - 0003-2409
DOI - 10.1111/j.1365-2044.1985.tb10663.x
Subject(s) - medicine , anesthesia , regional anaesthesia
Summary The design of and the results of laboratory performance tests on the Afya anaesthesia system are described. The system has a low inspiratory and expiratory resistance and the concentrations of ether delivered ranged from 75 to 130% of the vaporizer setting. The error was maximum at vaporizer settings of between 15 to 20%. The system is suitable for spontaneous and controlled ventilation of the lungs in both adults and infants.